| Power requirements: |
Supply voltage: |
24VDC ± 5V |
| |
Currents: |
Max.1A
|
| Resolution: |
|
3000 steps |
| Plunger drive: |
Principle: |
Quad encoder for loss of steps, detection of loss steps with
error message
|
| |
Travel: |
30mm |
| Plunger speed: |
|
1.2 sec. - 20 min./full stroke (speed ranges vary depending
on syringe size and tubing) |
| |
| Syringes: |
Type: |
XP: 50µl, 100µl, 250µl, 500µl,
1.0ml,
2.5ml, 5.0ml |
 |
Imprecision: |
0.05% CV within run at full
stroke (for 1000µl syringe and above) |
| Inaccuracy: |
<1.0% at full stroke |
| Fluid contact: |
Glass, PCTFE and Teflon |
| |
| Valve/valve drive: |
Valve options: |
120° 3-port distribution |
 |
Turn time: |
250ms between adjacent ports |
| Drive: |
Stepper motor with optical sensor for position detection
|
| Positions: |
Input, output, bypass
|
| Materials: |
Plug: virgin teflon
Body: PCTFE |
| Fittings: |
1/4-28"-UNF screw fitting for syringe |
| |
| Interface: |
Type: |
RS 232, RS 485 or CAN bus |
| |
Baudrate: |
9.600 or 38.400, 100K or 125K (CAN bus)
|
| |
Format: |
Data bits: 8 parity: none
Stop bits: 1 half duplex |
| |
Addressing: |
up to 15 XP/AL devices can be addressed individually |
| |
Communication: |
Data terminal and OEM protocol |
| Firmware: |
|
Programmable ramps, cutoffs, backlash compensation, plunger
speeds, delays and loops
Arbitrarily selectable plunger speed
Clearance compensation
Absolute and relative positions
Diagnostics
Flash-PROM |
| Environmental: |
|
Temperature: 17-37°C
Humidity: 10-85%
Non condensing RH at: 5°C-40°C |
| Dimensions: |
Height: |
127mm (+)1mm
|
| |
Width: |
45mm (+)1mm
|
| |
Depth: |
113mm |
| |
Weight: |
1.2kg |
| |
|
Home
/ Products
and Services
Impressum |